Publication

Influence of surface roughness on silicone rubber voice prostheses on in vitro biofilm formation and clinical lifetime in laryngectomised patients

Buijssen, K. J. D. A., Oosterhof, J. J. H., Basil, L., Waters, M., Duits, M. A., Busscher, H. J., van der Mei, H. C. & van der Laan, B. F. A. M., Dec-2017, In : Clinical Otolaryngology. 42, 6, p. 1235-1240 6 p.

Research output: Contribution to journalArticleAcademicpeer-review

Copy link to clipboard

Documents

  • Influence of surface roughness on silicone rubber voice prostheses on in vitro biofilm formation and clinical lifetime in laryngectomised patients

    Final publisher's version, 156 KB, PDF document

DOI

Objectives: Evaluation of the influence of a smooth surface moulding technique of silicone rubber indwelling voice prostheses on in vitro biofilm formation and analysis of the clinical in situ lifetime.

Design: Biofilm formation on smooth and Groningen ultra low resistance (URL) prostheses was studied in an artificial throat model. The clinical lifetime of smooth voice prostheses was compared to the previous lifetime of URL by counting the number of replacements in a consecutive 6-month period in the same patient. Participants: Eleven laryngectomised patients in follow-up who required frequent replacement of their voice prostheses.

Settings: Tertiary University Medical Center.

Results: Use of a smoother mould and less viscous silicone rubber yielded a decrease in surface roughness from 46 to 8 nm and was accompanied by a 40% reduction in the prevalence of bacteria and yeast in in vitro formed biofilms. Clinically, the lifetime was significantly (P

Conclusions: This combined in vitro and clinical study suggests that the choice of material and in particular its surface finishing may be determining factors with respect to the clinical lifetime of silicone rubber implants and devices failing due to biofilm formation.

Original languageEnglish
Pages (from-to)1235-1240
Number of pages6
JournalClinical Otolaryngology
Volume42
Issue number6
Publication statusPublished - Dec-2017

    Keywords

  • biofilm, laryngectomy, silicone, surface modification, surface roughness, TOPOGRAPHY, RESISTANCE, ADHESION

ID: 40129431