Tuncay, V., Zijlstra, J., Oudkerk, M. & van Ooijen, P. M. A., 10-Aug-2020, In : JOURNAL OF DIGITAL IMAGING. 33, 5, p. 1301-1305 5 p.

Research output: Contribution to journalArticleAcademicpeer-review

Copy link to clipboard

Documents

  • Design, Implementation, and Validation of a Pulsatile Heart Phantom Pump

    Final publisher's version, 556 KB, PDF document

    Request copy

DOI

The developments in Computed Tomography (CT) and Magnetic Resonance allow visualization of blood flow in vivo using these techniques. However, validation tests are needed to determine a gold standard. For the validation tests, controllable systems that can generate pulsatile flow are needed. In this study, we aimed to develop an affordable pulsatile pump and an artificial circulatory system to simulate the blood flow for validation purposes. Initially, the prerequisites for the phantom were pulsating flow output equal to that of the human cardiac pulse pattern; the flow pattern of the mimicked cardiac output should be equal to that of a human, a variable stroke volume (40-120 ml/beat), and a variable heart rate (60-170 bpm). The developed phantom setup was tested with CT scanner. A washout profile was created based on the image intensity of the selected slice. The test was successful for a heart rate of 70 bpm and a stroke volume of 68 ml, but the system failed to work at various heartbeats and stroke volumes. This was due to the problems with software of the microcontroller. As conclusion in this study, we present a proof of concept for a pulsatile heart phantom pump that can be used in validation tests.

Original languageEnglish
Pages (from-to)1301-1305
Number of pages5
JournalJOURNAL OF DIGITAL IMAGING
Volume33
Issue number5
Publication statusPublished - 10-Aug-2020

    Keywords

  • Pulsatile pump, Phantom design, Cardiovascular circulation mimicking, Computed Tomography imaging, Pulsatile flow